Three point temperature calibrations.
Calibration temperature range
Least uncertainty of calibrations (Best measurement)
For loggers without probes fitted (Internal sensor)
-15ºC to 40°C
± 0.061°C (routinely ±0.2°C from -15 to 40°C) in the F3 machine.
For loggers with probes fitted (External sensor)
-24°C to 130°C
±0.056°C (routinely ±0.2°C from -20 to 50°C) in dry block calibrator.
±0.036°C (routinely ±0.2°C from -20 to 50°C) in stirred liquid baths.
Single point temperature calibrations (new) and re-calibrations to loggers
Calibration temperature range
Least uncertainty of calibrations ( Best measurement )
For loggers without probes fitted (Internal sensor)
-24°C to 80°C
±0.061°C in stirred liquid baths with comm block fitted.
For loggers with probes fitted (External sensor)
-24°C to 130°C
±0.056°C At single temperature points in dry block calibrator.
±0.036°C At single temperature points in stirred liquid baths
Relative Humidity calibrations (new) and re-calibrations to loggers
Calibration humidity range
Least uncertainty of calibrations (Best measurement)
Single humidity points between 12% and 85% RH
12% to 85% RH
±1.0% RH at a dry bulb temperature between 20°C and 30°C and a dew point range of -6°C to 23°C
Calibration over the entire range from 12% to 85% RH
3.0% RH at a dry bulb temperature between 20°C and 30°C and a dew point range of -6°C to 23°C